ECECS 349

Measurements Lab


Supplementary Notes

 

Accurate Measurements with an oscilloscope.

 

An oscilloscope usually has an equivalent input impedance of 1 MW resistance in parallel with a capacitance of 10-20 pF. This high impedance is acceptable for DC measurements since the effects of the capacitance can be neglected and the high input resistance does not load the circuit to be measured provided that voltage across a circuit to be measured has low output resistance. On the other hand for transient circuits and a high frequency circuits, the input impedance of the probe plays a very important role and should be properly compensated for accurate measurements.

 

Figure 1 below shows the schematic representation of the input impedance of an oscilloscope. Resistance Rsc is in parallel with a small capacitor Csc. In general, in order to measure a voltage across the output terminals of a circuit, a coaxial cable is used to connect the scope to this circuit. Unfortunately, the coaxial cable introduces depending on the length of the cable, introduces relatively a large capacitance in parallel with the scope impedance which degrades the shape of the measured voltage further. In order to eliminate the effects the scope plus the cable impedance, a compensation circuit has to be introduced at the input of the coaxial line in series with the inner central conductor of the coaxial cable. The coaxial line with the compensation circuit is referred to as an “oscilloscope probe.”

 

Figure 1

 

The equivalent circuit of the complete measuring system is shown in Figure 1. The input of the scope is represented by Rsc in parallel with Csc. The coaxial line is approximated by the capacitance by Ccable. The probe is represented by a resistance Rp in parallel with a capacitance Cp.  The equivalent circuit is simplified as shown in Figure 2.

Figure 2.  Simplified circuit.

 

Figure 2 is further simplified as shown in Figure 3 where notations for the resistors and capacitors have been simplified

Figure 3. Further simplification of the circuit.

Here: R1=Rp, C1=Cp, R2=Rsc and C2=Csc+Ccable.

 

As can be seen from Figure 3 that the circuit is a simple voltage divided and the voltage Vo at the input of the scope can be writes as


                                                (Eq.1)

 

Here Z1 and Z2 are given by

 

 

Substituting these into Eq.1,

 

 

 


                                                                        (Eq.2)

If we choose R1 and C1 as:

 


Substituting these back into Eq. 2


Simplifying this equation, we obtain


 

Note that the result is equivalent to a simple voltage divider with two resistors in series with an output voltage at the input of the scope 1/10’th of the voltage applied to the probe tips. This is the 10X setting and independent of frequency (no s-dependence).

 

What would be the values of R1 and C1 if the probe were set at 1X setting?